DGSI EZ-DAQ Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 25
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
SOIL-DAQ™
CALIBRATION
Operator’s Manual EZ-DAQ™ version 1.4 23 of 25
Manually Adjusting Calibration Data.
The system allows manual input of A/D readings for re-entering calibration data or for
tweaking calibration data points to achieve proper calibration results.
2.0 Important Information
The above calibration steps have been provided for a yearly calibration check on the
DGSI EZ-DAQ touch screen.
Sensor types have been downloaded to the system by the factory before it shipped.
For example, sensor 1 is a 4-20 mA signal. This info has been coded into the touch
screen at the factory. A 200 mV/v sensor can’t be interchanged into this slot.
If a different sensor type is required, then factory recalibration is required.
Press here
for keypad.
Enter new desired A/D
number from keypad.
Press here
for keypad.
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25

Comentarios a estos manuales

Sin comentarios